 |
|
Volume 3, Number 3
|
|
July 2004 |
 |
FYI
Practical Contact Ultrasonics
- IIW Angle Beam Inspection |
by
Jim Houf*
Figures 1-3
Figures 4-6
The most critical
task prior to starting a contact angle beam ultrasonic inspection is
the UT operator’s calibration of the UT equipment with respect to the
specific part to be inspected. The term calibration in UT is
applied to both annual scope calibration and on site or field
calibration. Annual calibration determines that the scope meets horizontal
and vertical linearity requirements and can be done in house by a Level
III or externally by a qualified UT equipment supplier. On site calibration
using the IIW block or derivatives, the focus of this article, is performed
by the UT operator at the time of the inspection and consists of setting
up the scope presentation and sensitivity of the inspection unit to
perform a specific inspection.
The two most commonly used
UT calibration methods in the US are based on either the International
Institute of Welding (IIW) calibration block or a derivative that the
American Society of Mechanical Engineers' Boiler and Pressure Vessel
Code (ASME Code) refers to as a basic calibration block.
The IIW Block
IIW calibration blocks are
12 x 4 x 1 in. in size and are made of the same or acoustically
similar material as the part to be inspected. Predominant features of an
IIW block, Type 1 (Fig. 1), are two side drilled holes that are 0.06 and
2 in. in diameter and two notches. The first notch is a 0.06 in. deep
curved notch with an inner radius 1 in. from the marked reference point
and another 0.08 in. full width notch directly opposite the reference
point. One end of the IIW block is cut to a 4 in. radius from the
reference point. For illustration purposes, additional etched markings
located on the long edge of the block have been shown on the same side
or front of the block in Fig. 1.
The IIW block is designed to
permit the operator to perform multiple functions for both straight and
angle beam testing including distance and sensitivity calibration and
wedge angle verification. For straight beam calibration, the transducer
can be placed on the 1 in. surface at the reference point and the screen
width can be set by using the 4 in. reflection from the opposite side.
Block thickness can be used for 1 in. reflections. Because the reference
point is directly opposite the 0.08 in. notch, resolution can also be
determined.
Wedge Angle Verification
For angle beam testing,
the wedge angle can be checked by placing the transducer on the 1 in.
surface at point A in Fig. 2. In this example, a 70 degree point (shown
on the side of the wedge) is placed over the 70 degree mark etched on
the side of the block. The transducer is then moved back and forth until
the return signal from the 2 in. diameter hole is maximized on the
cathode ray tube (CRT) screen. The actual refracted angle can be read by
determining where the exit point mark aligns with markings on the side
of the block. When using 45 or 60 degree probes, the operator starts
with the transducer over the corresponding mark on the block and checks
the wedge angle in the same manner. It should be noted that most codes
and specifications permit the wedge angle to vary slightly within ±2
degrees of the designated angle, but the the tolerance should be
verified before continuing. If the wedge angle is within tolerance, the
operator can proceed to distance calibration or setting screen width.
Distance Calibration
To set up a CRT screen
width that represents the proper distance for the part being tested, the
operator must determine the length of the sound path in that thickness
of material, as was described in the previous article. Once the length
of a full skip distance is calculated, the screen width can be set. In
the following example we will set up a 10 in. screen.
Prior to starting the
distance calibration, good operating procedure is to make sure the
electrical zero, or main bang, is at or just off the left edge of the
CRT screen. If not, it is possible that the operator will be working
with the second reflection, which makes it impossible to calibrate the
machine. A simple way to determine this is to dampen a finger with
couplant and rub the bottom of the transducer face. The resulting signal
can be set to the left side of the screen.
Once the operator is
comfortable that the main bang is in the right place, the transducer is
placed on the 1 in. block surface above the marked reference point
(placement B, Fig. 2) and is aimed at the end of the block with the 4
in. radius. Using the range and delay controls (may be named differently
on newer machines, see manual), maximize the signal from the 4 in.
radius and set the reflector signal at the fourth major graticule on the
CRT screen. Then turn the transducer around and maximize the return
signal from the curved notch with the 1 in. radius. Set that signal at
the first major graticule. For smaller diameter transducers, it may be
necessary to move the transducer to the side of the 1 in. surface to get
a good signal back from the radiused notch. Switching back and forth
between the two transducer positions, the operator should continue to
adjust the controls until both signals line up on the proper graticules.
When this is accomplished, screen width is set to 10 in. with each major
graticule representing 1 in. of sound path.
Sensitivity Calibration
Sensitivity calibration is
done to provide an inspection reference level based on the amplitude
(height) of a signal from a reflector of a known size. On the IIW block,
that signal is generated from the 0.06 in. side drilled hole. To set
sensitivity, the block is turned over and the transducer is placed on
the 1 in. surface inboard of the 0.06 in. side drilled hole (placement
C, Fig. 2). The transducer is moved back and forth until the signal from
the hole is maximized on the CRT screen. Using the gain control, the
signal amplitude is then adjusted so that the maximized signal is set at
80 percent of full screen height (FSH). The amount of gain in decibels
(dB) is recorded and this gain value becomes the reference level for
inspection. Note that 80 percent FSH is commonly used, but some codes
and specifications may require other FSH values.
Once the UT system has been
calibrated, the operator can increase the gain setting to the scanning
level (dB value) dictated by the governing code or specification and
perform the inspection.
Alternative Calibration
Blocks
While the IIW block is a
very good calibration block, its large size and heavy weight are
inconvenient when carried in the field or when working out of position
and up in the air. Several other calibration blocks have been designed
that are smaller and lighter in weight.
Distance
Sensitivity Calibration Block.
The most commonly used alternative block is the distance
sensitivity calibration (DSC) block (Fig.
3). The DSC block measures 4 x 1 x 2.5 in.
which is considerably smaller and lighter than the IIW block
and will fit into a pocket. This block has a flat scanning surface with
a 1 in. radius at one end and a 3 in. radius at the other end. The 3
in. end of the block has a machined 0.375 in. deep, 0.031 in. wide flat
bottomed notch with a 2.625 in. radius from the reference point. Manufactured
commercially, these blocks can be purchased from many UT suppliers.
When using a block that
has a radius on both ends, it is important to remember that most of a
sound beam will reflect from an interface, so when the sound reflects
from the 1 in. radius, most of the sound returning towards the
transducer will reflect from the scanning surface and travel down
towards the 3 in. radius. This sound will then return to the scanning
surface but since it hits the scanning surface at the wrong angle to
enter the probe , it reflects downward towards the 1 in. radius and then
returns to the transducer, creating a second signal. As a result, it is
important to note that the distance between the back wall signals is the
sum of the distances from the radii to the reference point and the
second back wall (and all others) will show on the screen at 4 in., the
sum of 3 + 1 in., after the preceding back wall.
To perform a distance
calibration for a 5 in. screen, the transducer is placed at the
reference point and aimed at the end with the 1 in. radius. The first
return signal is maximized and placed on the second major graticule. Two
additional signals should be on the screen; a short signal from the
notch and a taller signal from the 3 in. radius. The signal from the 3
in. radius should be placed on the tenth major graticule and using the
delay and range controls, the 1 in. and 3 in. signals should be adjusted
until both fall on the proper graticule. When this is done, the scope
should be set for a 5 in. screen. The signal locations are represented
by the green signals shown on the 5 in. screen (Fig.
4a). To confirm
calibration, the transducer is reversed and aimed at the 3 in. radius.
If calibration is correct, the first signal on the screen will be the
notch (just past the fifth graticule) and the next will be the 3 in.
radius signal (sixth graticule). These positions are shown by the
signals shown in purple on the 5 in. screen presentation (Fig.
4a). As
mentioned above, since the signal from the 1 in. radius occurs 4 in.
later, it would come up at 7 in., which cannot be seen on a 5 in.
screen.
To perform a distance
calibration for a 10 in. screen, the transducer is again placed at the
reference point and aimed at the end with the 1 in. radius, but this
time the first return signal is maximized and placed on the first major
graticule. Because the distance between back walls is the sum of the
radii, the operator should also see back wall signals at 5 in. and 9
in., with notch signals 0.375 in. before each back wall signal, shown in
green in the 10 in. screen presentation (Fig.
4b). Again, the delay and
range controls should be used to position the back wall signals in the
proper places. To verify calibration, the transducer is again reversed
and aimed at the 3 in. radius. The operator should then see back wall
signals at 3 in. and 7 in., with notch signals slightly before each back
wall signal (signals shown in purple on 10 in. screen presentation in
Fig. 4b).
In either calibration, if
the second set of signals (shown in purple) do not come up on the screen
where they should be, the operator should use the delay control to
determine that the main bang is where it should be (at or off left side
of screen). If not, and if a first back wall is at that location, move
the main bang to the left edge of the screen and start over.
Sensitivity calibration using
a DSC block is performed in the same manner for either a 5 or 10 in.
screen. Once screen width has been set, the transducer is aimed at the
3 in. radius and the signal from the notch is maximized and set to 80
percent full screen height (or as detailed in governing documents).
The gain setting for this signal amplitude is used as the reference
level for the inspections. When using a 10 in. screen, there will be
two notch signals, at screen locations of 2.625 in. and 5.25 in. The
2.625 in. signal should be set to 80 percent FSH (or as required), and
if the operator is permitted to do so, peaks of the two signals can
be connected to create a rudimentary distance amplitude
correction (DAC) curve.
There is a correlation
between the IIW block and the DSC block. The signal amplitude from the
notch of an accurate DSC block should be within ± 2 dB of the signal
created by the 0.06 in. side drilled hole in the IIW block, and this
should be checked at regular intervals.
Distance
Calibration Block.
The distance calibration (DC) block (Fig.
5a), often called a 1-2 block, is similar in shape to the DSC block
but has a 1 in. radius and a 2 in. radius. It does not have a notch
for sensitivity calibration. For this reason, distance calibrations
can be performed but the operator must carry a separate block to set
the sensitivity level of the equipment
Half
Round Distance Calibration Block.
Another DC block is the half round block (Fig.
5b). Like the 1-2 block, it can be used for distance calibration
but a separate sensitivity block is required. The advantages to the
half round block are that that they can be readily manufactured by any
machine shop and, by using various radii, very narrow screen widths
can be set up.
IIW
Hit Block. The
IIW hit block is a small portable sensitivity calibration block
that is a 1 in. thick piece of material representing the corner of the
IIW block that contains the 0.06 in. side drilled hole. The block is
4 x 2 x 1 in. with an 0.06 in. side drilled hole 0.6 in. down from the
1 in. scanning surface and 1.4 in. in from the end of the block (Fig.
6). By using the hit block and a DC block mentioned earlier, calibration
for both distance and sensitivity is achieved. Distance is done as mentioned
above, and sensitivity is done using the hit block in the same manner
as is done on a full size IIW block.
*Jim Houf is Senior Manager of
ASNT’s Technical Services Department and administers all ASNT certification
programs. (800) 222-2768 X212,
(614) 274-6899 fax, <jhouf@asnt.org>.
[ The NDT Technician ]
Copyright © 2010 by the American Society for Nondestructive Testing, Inc. ASNT is not responsible for the authenticity or accuracy of information herein. Published opinions and statements do not necessarily reflect the opinion of ASNT. Products or services that are advertised or mentioned do not carry the endorsement or recommendation of ASNT.
IRRSP, NDT Handbook, The NDT Technician and www.asnt.org are trademarks of the American Society for Nondestructive Testing, Inc. ACCP, ASNT, Level III Study Guide, Materials Evaluation, Nondestructive Testing Handbook, Research in Nondestructive Evaluation and RNDE are registered trademarks of the American Society for Nondestructive Testing, Inc. ASNT exists to create a safer world by promoting the profession and technologies of nondestructive testing.