| Volume 3, Number 1 |
|
January 2004 |
 |
FYI
Practical Contact Ultrasonics
- Straight Beam Testing |
by Jim
Houf* and Bill Svekric†
What do thickness testers, sonar and fish finders have in common? All
use sound to detect foreign objects in a sound carrying medium.
While fish finders and sonar are used in
water and the majority of NDT straight beam inspections are performed in
steel, all three applications can be calibrated to the degree of
accuracy necessary to perform the required tasks. All use a zero degree
(through-thickness) longitudinal wave sound beam to interrogate the part
being inspected (in the case of fish finders and sonar, the part being
inspected is water). All rely on the same principle — sound traveling
through a uniform medium will reflect from the interface of that medium
with a material having different acoustic characteristics.
Typical Applications
For NDT, the two most common applications of straight beam
inspection are thickness measurement and verification of material
quality. In both applications, the transducer is placed on the surface
of the part using a liquid or gel couplant to couple the transducer to
the part. Couplant allows the sound beam to cross the gap between the
transducer and part. (For purposes of this article, it is assumed proper
use of couplant is understood.) The sound beam then travels through the
test piece, reflecting from the backwall or any internal planar
reflector and returns to the transducer. If equipment is properly
calibrated, the distance that the sound has traveled to the backwall or
reflector will be displayed either on a digital readout and/or
oscilloscope or liquid crystal display screen. Newer
instruments may use either single or dual element
transducers.
For thickness testing performed using a digital thickness tester,
the unit is calibrated so that the distance the sound beam travels to
the backwall is displayed as a digital readout on a liquid crystal
display. A basic digital thickness testing instrument consists of a
small battery powered, handheld generating unit and a three to four foot
coaxial cable with an integral dual element, delay line
transducer that plugs into the unit. The generating unit has a digital
liquid crystal display showing thicknesses in thousandths of an inch.
Most units have a known thickness round steel block attached to the main
unit that is used to standardize the system prior to use.
Dual Element, Delay Line Transducers
The dual element, delay line transducer has two
crystals, one transmitter and one receiver set side-by-side at the back
of the transducer assembly (Fig. 1). The sound carrying material between
the front face and the crystals (the delay lines) is divided longitudinally
into two sections so that the transmitted sound travels down one side
of the divider and returns up the other side to the receiver. The distance
between the sending crystal and the front face (delay line) permits
the near field effect to occur internally instead of in the part
and also eliminates entry surface noise, permitting the inspection of
thin materials.
Data Logging Units
Many units have data logging features that allow the operator to retain
thickness readings for downloading to a word processor after inspections
are completed. This feature can speed the inspection process
significantly. One point of caution should be noted — periodic
downloading is strongly recommended. The greater the amount of stored
information, the greater the loss should the unit fail to work properly.
The operator should check manufacturer recommendations for battery life
and replacement procedures.
Standardizing Equipment Settings
To operate a basic unit, the transducer cable is plugged into the unit
and the unit is turned on. A drop of couplant is then placed on the
built-in block on the unit and the transducer is coupled to the block to
allow the unit system to standardize. Note this is not calibration,
but is done to standardize the system. Since the thickness of the
block is known, the unit software will adjust the digital readout to
accommodate any changes in transmission characteristics caused by
changes in the coaxial cable or transducer. Modern units may have a
hidden transducer behind the block that lets the unit automatically
adjust for transducer wear. If the reading from the block is not
correct, remove the transducer, reapply couplant and try again. The
reading must match the block thickness or the next step,
calibration, cannot be done accurately.
Dried couplant or debris on the block surface can sometimes cause an
inaccurate reading. In this case, clean the surface of the block and try
again. Consult the operator manual for other options if this does not
help.
Calibration
To calibrate the system once it has been standardized,
place the transducer on a step wedge of acoustically similar material
that covers the thickness range of the part to be tested. For the following
example, a step wedge with five steps from 0.100 - 0.500 in. in 0.100
in. increments is used. The transducer is first placed on the 0.500
in. step and the readout is adjusted to match that thickness. The transducer
is next placed on the 0.100 in. step and the zero control knob
is used to set the readout for that step to 0.100 in. (Control functions
vary from unit to unit and the appropriate operator manual should be
referenced for the correct adjustment procedure.) The operator should
then return to the 0.500 in. step and repeat the process until both
the 0.500 and 0.100 in. readings are accurate. As a last step, check
the readings from the 0.200, 0.300 and 0.400 in. steps. If the readout
for each step is accurate, the unit is properly calibrated and the fixed
markers or graticules on the screen directly below the baseline can
be used to read the thickness (Fig 2). A significant advantage of the
liquid crystal display screen presentation on a digital thickness tester
is that the waveform often shows small discontinuities not large enough
to cause the digital readout to change or that may only cause the readout
to flicker back and forth between thicknesses.
Digital thickness testers with wave presentations
are calibrated in a similar way, but there will be both a digital readout
and a waveform presentation on a small liquid crystal display similar
to that found on a full size flaw detector. As with a flaw detector,
the horizontal baseline represents the sound path to the reflecting
surface, with the left side being near the transducer and the right
side being farther away from the transducer. Permanent horizontal and
vertical gridlines called major graticules are superimposed on
the screen cover to allow the operator to determine the distance the
sound has traveled (Fig. 2). The major graticules typically have five
subdivisions or minor graticules to improve reading accuracy.
Performing Inspections
Once the unit is calibrated, inspection can begin. To take a thickness
reading, place the transducer on the test surface at the first
inspection point. This causes a thickness reading to appear on the
digital readout screen. If the display remains constant, manually record
the reading or press the save or store key to record that reading
digitally. Move to the next inspection point and repeat the process. If
the display reading flickers or fluctuates between thicknesses, the
reading should not be recorded.
A reading may flicker for several reasons. If the
transducer is not seated properly on the part due to surface roughness
or a slightly curved surface, moving or rotating the transducer slightly
to get a solid coupling to the part may correct the problem. On smaller
diameter pipes or tubes, the transducer may rock side-to-side allowing
the couplant under the edge of the transducer to cause higher or thicker
readings than actually exist. This problem can often be corrected by
making sure the divider in the transducer is oriented perpendicular
to the axis of the pipe so that the major point of contact will include
both sides of the transducer (Fig. 3).
Readings that flicker may be related to the part itself (rough or pitted
backwall, multiple nonmetallic inclusions, or laminations in the part).
If two reflectors such as the backwall and the edge of a lamination are
sending approximately the same amount of sound back to the transducer,
the display may alternate between two thicknesses creating an unstable
reading. In these instances, the digital unit with a waveform
presentation can easily demonstrate the situation that is occurring.
Doubling is a confusing condition that can
occur when doing thickness testing of materials that range in thickness
from 0.040 - 0.080 in. in thickness. The sound bounces up and down twice
in the part before detection by the crystal. The reading is twice or
double the actual thickness (Fig. 4). Doubling should be suspected when
the operator gets consistently thin readings that abruptly become almost
twice that of the adjoining area.
As an example, readings drop consistently to around 0.060 in. as the
operator moves the transducer across the part. Abruptly, the readings
double with the next reading taken as 0.115 in. and a succeeding reading
at 0.110 in. This indicates that doubling may be occurring. At this
point, the operator should substitute a transducer with crystals
manufactured at a slight angle (roof angle) to help aim the sound
beam more favorably.
When using a full flaw detector or scope, the operator must
select the type of transducer (single or dual element), the diameter,
and the frequency. Older scopes require the operator to match the
frequency setting on the unit to the transducer frequency. Newer units
may match the two automatically. When selecting transducers, keep in
mind that some frequency/diameter combinations can result in near fields
that are unacceptable for the proposed thickness range. If this is in
question, the operator should check with his Level III. One advantage
obtained by using a dual element transducer is the tendency of the unit
to trigger from the nearest reflector, such as internal corrosion
or pitting. Single element transducers tend to read the reflector with
the largest surface area. For example, a deep farside pit with a
small surface area may not be picked out from the larger backwall
signal. As a result, single element transducers may result in thicker
readings than what may actually exist at that point in the part or may
miss small reflectors as described in the example.
An advantage of using single element transducers is that coating
thicknesses may be eliminated from the readings. This is done by reading
the thickness value between the first and second backwall reflections.
The same result may be obtained between the second and third backwall
reflections if the display is difficult to read. Modern instruments have
electronic gates that may be set to automate this process. The fact that
coatings do not need to be removed saves time and eliminates the cost of
coating removal and reapplication.
When a full flaw detector or scope is used for straight beam testing,
setting the screen width and calibrating the instrument are one and the
same. To set screen width, the transducer is first placed on a piece of
acoustically similar material of known thickness such as a step wedge.
Using two reflectors at a known distance, the UT operator then uses the
range and delay controls to set screen width. The delay
control shifts the entire screen display right or left without changing
the distance between vertical traces. The range control expands or
contracts the distance between vertical traces. (Consult UT reference
text for technical explanation of range and delay.) These two control
functions interact and it is necessary to alternately adjust each to
obtain the desired screen presentation. Check the operator manual to
confirm which knobs or touch pad keys control the range and delay
functions for your machine.
To set up the screen width using a single element
transducer, the operator can get two backwall reflections from the 0.500
in. step on the screen and by spreading the screen with the range control
and shifting the display to the left with the delay, can set the left
trace (first backwall echo) on the number 5 graticule and the right
trace (second backwall echo) on the number 10 graticule. Once the two
traces are set this way, the full screen width is 1.0 in., with major
graticules representing 0.1 in. and the scope calibrated for straight
beam inspection from 0 - 1 in. (Fig. 5).
The operator should next use the gain control to set the
amplitude or trace height from the backwall echo from the
test part to 100 percent of full screen height. Full screen height is
used most commonly for backwall amplitude but operators should check
their NDT procedure or job specification to determine exact requirements
for the specific job.
Once the scope is calibrated and the amplitude set, the inspection
technique is the same as for digital thickness testers, with the
thickness being read from the horizontal graticules. When checking for
laminations or inclusions, the depth of these discontinuities can be
determined in the same manner. However, the governing specification may
require acceptance or rejection to be based on amplitude of the
resulting trace, the loss of backwall amplitude (loss of back)
method or a combination of these two.
Amplitude can be read directly from the liquid crystal display or
cathode ray tube screen, while full loss of back means the
backwall trace drops completely off the screen. The specification may
also call for a combination of the two methods and require rejection
only if a reflector has a certain amplitude and causes a certain
percentage of loss of backwall amplitude. An example of this would be a
40 percent full screen height reflector with a 50 percent loss of back.
Mapping
When a rejectable lamination or inclusion is found, the customer may
want the operator to map the rejectable area. The most common
method for mapping is to slide the transducer across the edge of the
area where the signal first appears until the screen presentation meets
the rejection criteria. The plate or part is then marked beside the
center of the transducer. The transducer is then moved laterally until
the same condition occurs and the plate is again marked. By repeating
this procedure, the area of the defect can be outlined and mapped on the
inspection report and on the surface of the piece being tested.
Conclusion
While this article is intended to provide information
for the beginning UT operator, it may not be sufficient in some situations.
If additional clarification is required, operators should consult their
Level III. TNT
|
| Figure 1. Schematic of transducer assembly. |
|
| Figure 2. Liquid crystal display or
cathode ray tube screen presentation for digital thickness tester
showing major graticules. |
|
| Figure 3. Dual element divider orientation
on tubing or piping with small diameters. |
|
| Figure 4. Diagram of sound path demonstrates
doubling phenomenon that can sometimes occur when testing material
thickness in the range of 0.04 - 0.08 in. |
|
| Figure 5. One inch liquid crystal display
or cathode ray tube screen presentation from 0.5 in. step wedge. |
*Jim Houf is Senior Manager of ASNT’s Technical
Services Department and is responsible for administration of all ASNT
certification programs. He’s been involved in NDT since 1972 and an ASNT
Level NDT Level III since 1984. He currently holds ASNT NDT and ACCP
Professional Level III certificates in four NDT methods. He’s also an
AWS Senior Welding Inspector and an ASQ Certified Quality Auditor. (800)
222-2768 X212, (614) 274-6899 fax, <jhouf@asnt.org>.
†Bill Svekric is President of Welding
Consultants, Inc., Columbus, Ohio. He earned a Bachelor's Degree in
Welding Engineering (BWE) from the Ohio State University. He received
his ASNT NDT Level III certificates in 1978 and currently holds ASNT NDT
and ACCP certification in four NDT methods. He is an AWS CWI and is an
ASNT Fellow. (614) 258-7018, (614) 258-1996 fax; <wciinc@aol.com>.
Copyright © 2012 by the American Society for Nondestructive Testing, Inc. ASNT is not responsible for the authenticity or accuracy of information herein. Published opinions and statements do not necessarily reflect the opinion of ASNT. Products or services that are advertised or mentioned do not carry the endorsement or recommendation of ASNT.
IRRSP, NDT Handbook, The NDT Technician and www.asnt.org are trademarks of the American Society for Nondestructive Testing, Inc. ACCP, ASNT, Level III Study Guide, Materials Evaluation, Nondestructive Testing Handbook, Research in Nondestructive Evaluation and RNDE are registered trademarks of the American Society for Nondestructive Testing, Inc. ASNT exists to create a safer world by promoting the profession and technologies of nondestructive testing.