|
Back to Basics
[ click
here for the Back to Basics Archive ]
Calibration for Nondestructive
Testing
by Stuart Kleven*,
Israel Vasquez† and David
Atkins‡
|
Often, technicians are required to verify a current calibration
sticker on an instrument and then to perform the NDT. Technicians
are exposed to the calibration process or specification requirements
only as they reference documents, and most training materials
do not adequately address calibration requirements and the effects
of out of calibration conditions. This article will provide technicians
with a closer look at the history, standards and requirements
behind the calibration sticker.
Bruce Crouse
Contributing Editor
|
INTRODUCTION
Standards are
as old as human society itself. During the era of the pharaohs, calibration
standards were used for measurement during the construction of the great
pyramids at Giza. The Royal Cubit was the length from the back of the
pharaoh's elbow to the tip of his middle finger. This was established
in a stable material such as granite. Then working standards were made
and compared to the master. These were issued to workers, and they were
required to return the working standards for comparison against the
Royal Cubit master every full moon. Failure to submit the working standard
for verification was punishable by death. The accuracy obtained by the
use of the working standards was phenomenal. In a structure measuring
approximately 230.4 m (756 ft), the Egyptian builders were accurate
to within 115 mm (4.5 in.). This is about a 0.05% accuracy (National
Council of Standards Laboratories, 2006). Weights (Figure 1) from most
ancient societies have been discovered and are accurate within tenths
or hundredths of a gram when compared to the required standard weight.
Like us, ancient people were very concerned with maintaining accuracy.
Even within the Bible, it was stated that "the Lord detests differing
weights, and dishonest scales do not please him" (Proverbs 20:23).
Over
the years, a number of calibration documents
have been specified for use.
Today, the calibration of testing equipment would appear
to be a subject that is unimportant or secondary to their immediate
purpose. Many companies merely send out equipment and indicate on purchasing
documents that a calibration certification is required. This ignores
the question of what a proper calibration is. One complicating factor
is the profusion of different types of equipment and equipment manufacturers,
which can also add confusion to the situation. For instance, some units
are analog and others are digital. What rules apply? Can one type be
used to verify the other? In addition, most governing specifications
in nondestructive testing only give the frequency that is expected for
the calibration of certain items, but rarely give any limits or tolerances.
This can cause quite a quandary for the inexperienced and even the experienced
quality professional. This paper will attempt to address many of the
issues that have caused problems and make possible suggestions for standardizing
some calibrations.
CALIBRATION DOCUMENTS
Over the years, a number of calibration documents have
been specified for use in NDT. The old standby for many years was MIL-C-45662A.
This was later converted to MIL-STD-45662 and eventually revised
to MIL-STD-45662A. This stood for quite a number of years as
the only guideline for calibration. Many auditors and customers simply
said that it was up to the calibration lab to establish a system and
then follow it based on the guidance of MIL-STD-45662A. In
1984, MIL-HDBK-52A was issued to provide guidance in applying
MIL-STD-45662A. In 1989, MIL-HDBK-52B was released
to further clarify the application of the military standard. Recent
developments in the calibration area have produced a number of documents
and schemes addressing calibration, both for the calibrating laboratory
and for the company specifying calibration. These include ANSI Z540-1
(the US equivalent to ISO Guide 25), ISO 10012-1,
and ISO 17025 (formerly ISO Guide 25).

Figure 1 — Ancient stone weights.
ISO 17025, General Requirements for the Competence
of Testing and Calibration Laboratories, is the main document used
today. It consists of the following components or characteristics:
- scope
- normative references
- terms and definitions
- management requirements (includes ISO 9000 components)
- technical requirements (personnel, equipment, traceability, lab
conditions, QA reports, and so forth)
- annexes.
CALIBRATION SOURCES
The first step in calibration is to find a dependable
calibration source for performing the work. The typical approach is
to determine who can supply the calibration at the lowest price. Taking
this course could, however, prove to be a problem later. The first step
to finding a dependable source is to look for some sort of nationally
recognized acceptance (like accreditation). This means that someone
from an independent, third-party source has assessed the calibration
source and found it compliant to the standard or specification. While
this does not necessarily ensure perfection, it removes quite a few
factors that can affect the calibration, such as the use of adequate
quality procedures, valid test methods, calibrated equipment with traceability
to the National Institute of Standards and Technology (NIST), an understanding
of the accuracy and uncertainty limits, and proper personnel training
and qualification. Laboratories with accreditation to the National Voluntary
Laboratory Accreditation Program (NVLAP), the American Association for
Laboratory Accreditation (A2LA), the Laboratory Accreditation Bureau
(LAB) or other accredited registrars are preferred.
Each of these accrediting bodies maintains a listing of
the calibration laboratories that have been assessed. Assessment may
be performed by the company requiring the calibration as well. This
may be achieved by a desk audit (filling out a checklist), an onsite
audit or by inspecting and reviewing the calibration certificate and
the instrument upon receipt. Other factors can affect the selection
of calibration laboratories as well. If a delicate or bulky equipment
requires calibration, shipment across the country can cause problems.
The length of time a piece of equipment is out for calibration may necessitate
the acquisition of two units to maintain calibrated status. In addition,
the handling in shipping can cause inherent troubles, possibly by damaging
the unit during transit or by affecting the newly calibrated condition.
The lack of personal contact and confidence is missing when the calibrating
laboratory cannot be readily visited. Onsite audits are less likely
to be performed when the laboratory is not in close proximity to the
company using their services.
DEFINITIONS
It would be helpful to define terms with regard to calibration.
The following definitions were obtained from the documents previously
cited in this paper.
Calibration
The set of operations which establish, under specified
conditions, the relationship between values indicated by a measuring
instrument or measuring system, and the corresponding standard or known
values derived from the standard. In simple terms, comparing a known
against an unknown. (Not the same as standardization, which is sometimes
called calibration. For example, radiographic film densitometers are
standardized with a "working" step tablet, and calibrated
with an NIST traceable "master" step tablet.) Calibration
can refer to several different types of actions taken by a company:
outside calibration of master equipment directly traceable to NIST;
in-house calibration of equipment to an in-house master directly calibrated
to NIST; in-house calibration of equipment by an outside source with
master equipment traceable to NIST.
Accuracy
Conformity to a certified or approved standard. A measure
of closeness of agreement between a measured result and the true value.
Accuracy is a qualitative concept.
Uncertainty
The result of the evaluation aimed at characterizing the
range within which the true value of a measurand is estimated to lie,
generally within a given likelihood.
Under MIL-STD-45662, collective uncertainty shall
not exceed 25% of the acceptable tolerance for the characteristic being
calibrated. This is commonly called the 4:1 test accuracy ratio. Sources
contributing to uncertainty include the reference standard or materials
used, the method and equipment used, environmental conditions, the condition
of the item being calibrated, and the technician performing the calibration.
For ISO 17025, the uncertainty must be developed
and justified based on analysis and consideration of the conditions
listed above. Some customer specifications have listed uncertainty requirements
ranging from 4:1 to 10:1. Care should be exercised in agreeing to these
limits, since they may be unattainable under certain situations.
Tolerance
The set of values for a measurand for which the error
of a measuring instrument is intended to lie within specified limits
(allowable deviation from a certified or approved standard, such as
+/- 10%).
Stability
The ability of a measuring instrument to constantly maintain
its metrological characteristics.
Drift
The slow variation with time of a metrological characteristic
of a measuring instrument (especially applicable to electronic instrumentation).
Traceability
The ability to relate individual measurement results through
an unbroken chain of calibrations to one of the following:
- US National Standard maintained by NIST, or other country standard
correlated to the US National Standard
- fundamental or physical constants with values assigned or accepted
by NIST
- ratio type calibrations
- comparison to consensus standards.
Adequacy of the Measurement Standard
The measurement standard shall be traceable and have the
accuracy, stability, range and resolution required for the intended
use.
Precision
This is the closeness of multiple readings to each other.
Bias
This is the accuracy of the closeness of multiple readings
to the true value.
STANDARDS
Calibration standards are normally established in the
US by NIST, formerly known as the National Bureau of Standards (NBS).
Each country has their own respective agency. For a complete listing,
visit NIST's Web site at www.nist.gov/oiaa/national.htm.
Traceablility to these known standards is the normal method used for
calibration. At times, NIST has established physical constants that
are used for calibration, such as the speed of light in a vacuum, the
Josephson frequency voltage ratio, and quantized Hall resistance. Consensus
standards may be established where no national standards exist. These
are standards that everyone agrees on as a standardized means for calibration.
If none of the above exists, original equipment manufacturer (OEM) standards
may be used.
FREQUENCY OF CALIBRATION
While some specifications or standards do establish a
calibration frequency, most leave it up to the user. The natural inclination
is to set the frequency as long as possible to reduce calibration costs.
This of course may increase the likelihood that, should the instrument
be received by the calibration source with a "significant out-of-tolerance"
condition, all products affected by this instrument over the extended
period of use is now suspect. The proper way to establish the correct
frequency is to examine the literature from the equipment manufacturer
to determine the stability of the unit. The next step is to set a shorter
frequency based on the stability, purpose and usage of the equipment.
Typically, electronic instruments tend to drift or get out of calibration
sooner than fixed gages or measuring tools such as microthickness gages.
To give an example, an electronic ohmmeter may be set at a quarterly
frequency if it is used fairly often. Then, after four cycles where
documentation demonstrates that the equipment has maintained its accuracy
throughout the calibration intervals, the frequency could be extended
to six months. This could be repeated for three or four cycles and again,
if no change is noted, a slightly longer frequency could be established,
such as annually. If an instrument comes back from calibration with
an out-of-tolerance condition, it may be advisable to shorten the frequency
for at least one cycle to determine whether or not the condition is
due to inherent problems with the instrument or if it was an isolated
incident. (See "Out-of-Tolerance Conditions," below, for more
information.)
Several newer standards have determined ways to set the
frequency. The following guidelines for the determination of confirmation
intervals for measuring equipment are described in detail in ISO
10012-1, Annex A:
- automatic or staircase method
- control chart method
- calendar time method
- "in-use" time method
- "in-service," or black box method.
UNCERTAINTY
Uncertainty is the amount of deviation allowed within
the tolerance. Based on the current ISO 17025, the uncertainty
must be stated by the calibration laboratory. Each laboratory shall
have and apply a procedure to estimate the uncertainty of measurement
for all types of calibrations. They are required to develop an uncertainty
budget based on internal factors, such as identifying all the components
of uncertainty and making the best possible estimation, and ensuring
that the form of reporting does not give an exaggerated impression of
accuracy. These factors include environment, physical constants, the
measuring procedure, definition of the measurand, the measuring object,
the metrologist or technician, software and calculations, measurement
setup, the actual measuring equipment, and the reference element. Once
each factor is identified as providing uncertainty, the amount must
be determined to the best possible degree. All these factors must then
be added to determine the total uncertainty. The process is further
refined by determining which factors add the most uncertainty and which
add cost that may be unreasonable when attempting to perform a calibration.
As mentioned above, under MIL-STD-45662A, it is usually 25%
of the tolerance and is commonly called the 4:1 test accuracy ratio.
While this is the goal, sometimes it is difficult to find anyone who
can hold the 25%, and it may be necessary to accept lower ratios.
In addition, the accuracy of the calibration can be affected
by the number of times the instrument under calibration is removed from
the NIST-traceable master. For instance, if a company sends their light
meter out to a company that had sent their meter out and their source
is three or four times removed from NIST, the uncertainty may be larger
than the allowable tolerance. The uncertainty for each time removed
from NIST must be established. This is where many companies make the
mistake of thinking that they are getting a valid NIST-traceable calibration,
when in fact many calibration sources do not check the number of times
removed or the uncertainty involved. This must be specified, since multiple
sources may be involved. The use of analog meters or instruments to
calibrate digital equipment should not be permitted. Analog equipment
requires interpolation in between lines to attempt to obtain a reading.
This is not accurate enough to calibrate digital equipment. The reverse,
however, is permissible. The readout of digital equipment is three or
four decimal places. This would allow for an accurate observation of
units on an analog unit that cannot be read with digital accuracy.
WRITING PURCHASE ORDERS FOR CALIBRATION SERVICES
A written purchase order communicates all the requirements
for calibration to the calibration source. Verbal purchase orders should
be avoided entirely since they can leave too much open for interpretation
and can cause problems later on if an instrument does not function within
a specific range as expected. Written purchase orders should include
the following items:
- the standard or controlling document (for example, ANSI Z540-1,
ISO-17025 or MIL-STD-45662A)
- the piece of equipment, identified by model number, type and serial
number
- the required tolerance and requirement to record uncertainty
- the required frequency for calibration
- the requirement to indicate the "as- received" and "as-left"
condition
- any environmental conditions that must be recorded (such as temperature
or humidity)
- any traceability requirements, such as traceability to NIST, and
date of calibration of master equipment, model number and serial number
- a requirement for the signature of the person performing the calibration,
date of calibration and date due
- a requirement for recording the procedure number used for performing
the calibration, as applicable
- a requirement for a label or sticker to be applied, giving the
name of the person performing the calibration, date performed and
date due (also a requirement for the application of a tamper-proof
seal to prevent readjustment of equipment, if applicable).
A blanket purchase order is an acceptable means of conveying
this information. This way, the company does not have to develop a purchase
order each time instruments are sent out. The blanket purchase order
should have an expiration date, and also a review date (at least annually)
for compliance. The company sending items out for calibration should
provide a complete list of all nondestructive testing equipment or instruments
to the calibration laboratory (ether internal or external), referencing
the names, nomenclature, serial number or asset number, intervals of
calibration, ranges of use, intended use, and the specification and/or
manufacturer's tolerances. This will prevent noncompliances during subsequent
audits.
VERIFYING STATUS OF EQUIPMENT AFTER CALIBRATION
After a piece of equipment is returned, the status of
the calibration must be verified. This is basically a receiving inspection
that is performed. All of the purchase order requirements listed above
should be checked to make sure the calibration source recorded all the
information requested. Experience gained by the authors over a number
of years dictates that the test results should be reviewed, even though
the calibration source has indicated acceptability. It is easy for a
decimal point to be misplaced or information recorded that is out of
tolerance and was missed during completion and final review of the certificate
of calibration. This will prevent acceptance of data that is out of
tolerance and that could possibly be discovered by an auditor later.
If the instrument was actually sent back out of tolerance, then all
hardware accepted during nondestructive tests using that instrument
could be in question and could cause severe problems and expense, such
as grounding aircraft, disassembly and retesting. Review of the dates
of calibration of master equipment traceable to NIST is also important.
It is not uncommon to find calibration sources that do not send their
master equipment out on a frequent basis. For example, one calibration
source listed the NIST number for their master equipment. When asked
for the date of calibration, it was twenty years ago! So just having
a NIST number is not enough.
There are also other methods for checking incoming calibrated
equipment. If the instrument has a battery, turn the instrument on to
see if the unit works. Sometimes placing the unit in a box or container
for shipping after calibration can accidentally activate the "on"
button. The unit would then be dead "as-received." Another
easy check is to examine the box as received to see if any physical
damage had occurred during shipping. If a delicate instrument was inside,
the calibration could be affected. If a similar or like piece of equipment
is available, compare the two side by side. For instance, if the ambient
light in a penetrant testing booth is zero on one calibrated light meter
and 32 lux (3 ftc) on another, there is a problem with one of the two
meters, possibly the one just received from calibration. Another easy
check for light meters is to turn on the unit and cover the sensor.
It should read zero if all light is excluded.
A calibration review checklist encompassing the ten items
listed under "Writing Purchase Orders," above, is a viable
way to assure that the certificates of calibration are complete and
provide the correct information. This checklist can then be attached
to the certificate as documented evidence of review. See Table 1 for
an example of a checklist.

Table 1 — Sample calibration certification review record.
OUT-OF-TOLERANCE CONDITIONS
When an instrument is received with an out-of-tolerance
condition by the calibration source, a determination must be made concerning
items that were accepted using equipment that was checked using the
out-of-tolerance instrument. An impact assessment must be performed
to determine if the out-of-tolerance condition is significant and adversely
affects product quality, measurement integrity and/or safety. For example,
let's say an ultrasonic light meter was sent out for calibration at
+/- 5% of a standard, and it was reported that it was out of calibration
by +1%. If the light reading is 20 µW/mm2
(12.9 mW/in.2), 6% of that measurement
would be 1.2 µW/mm2
(774 µW/in.2). If a
further reading was taken after a few days and it was 15 µW/mm2
(9.7 mW/in.2), and then one subtracted
1.2 µW/mm2 (774 µW/in.2)
from that reading, the result would be 13.8 µW/mm2
(8.9 mW/in.2). This is still above
a 12 µW/mm2 (7.7 mW/in.2)
minimum reading some primary contractors require. That would mean that
the out-of-tolerance condition did not have a significant effect on
quality. This analysis would need to be written up and placed with the
calibration data to demonstrate an acceptable condition despite the
fact that the instrument was received in an out-of-tolerance condition.
CONCLUSION
While calibration may seem to be of minor importance,
it can make or break the validity of a test or test system. Proper attention
to the requirements for calibration can help reduce the possibility
that this would contribute to poor or improper NDT. Of course, normal
operations could also affect test equipment at any time during operation.
Other factors such as age, misuse or amount of use could cause premature
drift or out-of-tolerance conditions. Correct application of the specifications
will facilitate better control of the calibration system and related
equipment. Proper training of personnel to recognize problems and take
preventative action (as opposed to having to take corrective action)
is another part of the equation for a successful calibration system.
REFERENCES
ANSI Z540-1: General Requirements for Calibration
Laboratories and Measuring and Test Equipment, Washington, DC,
American National Standards Institute, 1994.
ISO 9000: Quality Management Systems — Fundamentals
and Vocabulary, Geneva, International Organization for Standardization,
2005.
ISO 10012-1: Measurement Management Systems —
Requirements for Measurement Processes and Measuring Equipment,
Geneva, International Organization for Standardization, 2003.
ISO 17025, General Requirements for the Competence
of Testing and Calibration Laboratories, Geneva, International
Organization for Standardization, 2005.
ISO Guide 25: General Requirements for the Competence
of Calibration and Testing Laboratories, Geneva, International
Organization for Standardization, 1990.
MIL-C-45662A: Calibration Systems Requirements,
Washington, DC, US Department of Defense, 1962.
MIL-HDBK-52A: Evaluation of Contractor's Calibration
System, Washington, DC, US Department of Defense, 1984.
MIL-HDBK-52B: Evaluation of Contractor's Calibration
System, Washington, DC, US Department of Defense, 1989.
MIL-STD-45662: Calibration Systems Requirements,
Washington, DC, US Department of Defense, 1980.
MIL-STD-45662A: Calibration Systems Requirements,
Washington, DC, US Department of Defense, 1995.
National Council of Standards Laboratories, "Papyrus
Story," NCSL Web site, available at www.ncsli.org/misc/cubit.cfm,
accessed 2006.
* Alloyweld Inspection
Company, Inc., 796 Maple Lane, Bensenville, IL 60106; (630) 595-2145;
fax (630) 595-2128; e-mail skleven@alloyweldinspection.com.
†
Vastek Consulting, 7290 Jasmine Dr., Hanover Park, IL 60103; (630) 213-3432;
fax (630) 213-3495; e-mail vastek@aol.com.
‡
Packer Engineering, 1950 N. Washington St., Naperville, IL 60563; (630)
505-5722; fax (630) 505-1986; e-mail datkins@packereng.com.
Copyright ©
2006 by the American Society for Nondestructive Testing, Inc. All rights
reserved.
[ Back
to Materials Evaluation ]
Copyright © 2012 by the American Society for Nondestructive Testing, Inc. ASNT is not responsible for the authenticity or accuracy of information herein. Published opinions and statements do not necessarily reflect the opinion of ASNT. Products or services that are advertised or mentioned do not carry the endorsement or recommendation of ASNT.
IRRSP, NDT Handbook, The NDT Technician and www.asnt.org are trademarks of the American Society for Nondestructive Testing, Inc. ACCP, ASNT, Level III Study Guide, Materials Evaluation, Nondestructive Testing Handbook, Research in Nondestructive Evaluation and RNDE are registered trademarks of the American Society for Nondestructive Testing, Inc. ASNT exists to create a safer world by promoting the profession and technologies of nondestructive testing.
|