Probes for Remote
Field Testing of Heat
Exchangers: Configurations and Capabilities
by David
Mackintosh* and Brian Beresford+
Figure 1-4
Table
1-2
Principle of Operation
Remote
field testing is a technique commonly used for NDT of small bore steel
tubes such as those found in heat exchangers. The term remote field
derives from the relatively large spacing of two to three tube
diameters between the transmitting and receiving coils. The large
spacing ensures that there is no direct coupling between the coils.
Instead, the magnetic field travels from the transmitting (or exciter)
coil outwards through the tube wall, axially along the outside of the
tube and back through the tube wall to the receiving (detector) coil (Figure
1). Remote field testing is therefore known as a through
transmission technique. When the probe moves into an area with wall
thinning, the shielding effect of the tube wall is reduced and the
field at the detector coil changes in two ways: its strength increases
and its time of flight decreases. Remote field testing has
approximately equal sensitivity to metal loss on the inside and
outside of the tube wall. Metal loss increases the amplitude; it does
not matter whether that increase occurs on the inside or outside wall.
The overall strengths and limitations of remote field testing are
described in detail in ASTM E 2096: Standard Practice for In Situ
Examination of Ferromagnetic Heat-exchanger Tubes Using Remote Field
Testing (ASTM International, 2000).
Detector and Exciter Response
For small bore tubes, the exciter coil is usually made larger than the
detector (Figure 1) to produce
a higher magnetic field and therefore higher sensitivity to pits. The
detector coil size is based on a trade off between two values: the
number of turns (the more turns, the greater the voltage output) and
the dimensions (the slimmer the coil, the higher the resolution and
sensitivity). At a small area of metal loss, the basic probe in Figure
1 actually produces two indications, one for each coil. An area of
metal loss long enough to cover both coils doubles the response. The
remote field testing inspector must be aware of these effects to
obtain accurate results.
Remote
field testing has approximately equal sensitivity to metal loss
on the inside and outside of the tube wall.
Detector Configurations
The single bobbin coil is known as an absolute coil because its
voltage output is directly proportional to the local value of the
magnetic flux density (Figure 2).
Absolute coils are excellent for sizing large volume discontinuities
such as general wall loss and erosion. A differential detector is
formed by wiring two adjacent absolute coils with opposing outputs.
The differential detector subtracts out slow variations in magnetic
flux density and produces a larger response to sharp changes in the
field, such as those at pits. Differential coils can also be used to
size large volume metal loss, although with less accuracy than the
absolute coil.
Both absolute and differential bobbin coils can be split into a set
of smaller coils, making what is known as an array detector (Figure
2). Each coil in the array is sensitive to the immediately
adjacent segment of the tube circumference. Array probes tend to be
more sensitive to small discontinuities and they also produce data
that intuitively indicates the circumferential extent of metal loss
from the response of each coil. The use of array detectors is becoming
more common and likely represents the next step up in accuracy and
sensitivity for remote field testing. One of the great promises of
array detectors is improved performance at tube support plates. The
main disadvantage of arrays is that they produce more data for the
analyst to interpret, a problem which can be offset by software with
improved capabilities for displaying and manipulating multichannel
data.
The configurations discussed in this section are summed up in Table
1. The five point scale is used for clarity, but is based on
personal experience and is not meant to imply a quantitative
comparison.
Basic Probe at Tube Support Plates
Discontinuities in heat exchangers often occur near tube support
plates and baffle plates. (Baffle plates serve a dual function of
supporting the tubes and directing the flow of fluid within the heat
exchanger. The discussion of tube support plates in this paper applies
also to baffle plates.) Near tube support plate areas, remote field
testing has limited sensitivity and accuracy due to the interruption
by the tube support plate of the field traveling on the outside of the
tube. Figure 3 shows that on
one side of the tube support plate, the basic remote field testing
probe creates what could be called a shadow, which is a zone scanned
by the exciter but not the detector. Mainly because of its size, the
exciter coil is less sensitive to pits than the detector, so the
shadow is an area of limited sensitivity to pits. One way to remove
the shadow is to use a probe with more coils, as explained in the next
section.
Multiple Coil Probes
The need to detect and size disconti-nuities near tube support plates
has led to the use of probes that have more than just one exciter and
one detector (Figure 4). Adding
an extra exciter to a probe (Figure
4b) improves performance near tube support plates by removing the
shadow zone and gives excellent sensitivity to pits. The disadvantage
of this double exciter probe is the complexity of its indications. For
example, one small area of metal loss may produce three indications,
one for each coil. The double exciter also creates interference
between long metal loss indications that are sensed by the three coils
simultaneously, which reduces the accuracy of sizing large volume
metal loss or pits near or inside large volume metal loss. The double
detector probe (Figure 4c)
creates only two indications in each detector channel for a small area
of metal loss (similar to the basic probe in Figure
1), with correspondingly less interference between different areas
of metal loss. The disadvantage of the double detector configuration
is that there is double the data to display and analyze, and the
indications from the two detectors are not aligned in the strip
charts.
The configurations discussed in this section are summed up in Table
2. Again, the five point scale is used for clarity, but is based
on personal experience and is not meant to imply a quantitative
comparison.
Conclusion
There is no single remote field testing probe or detector that can
detect and size all types of metal loss in all regions of a heat
exchanger with maximum effectiveness. The strengths and limitations of
each probe configuration must be understood to improve data analysis
and to choose the best probe for the application, tube discontinuities
and priorities of the plant owner. Plant owners often choose the
remote field test vendor who demonstrates the best sensitivity to
pits, but there is usually a trade off between performance for pits
and for large volume metal loss.
References
ASTM International, ASTM E 2096: Standard Practice for In Situ
Examination of Ferromagnetic Heat-exchanger Tubes Using Remote Field
Testing, West Conshohocken, Pennsylvania, ASTM International,
2000.
*
Canspec Group, Inc., 7450 - 18 St. NW, Edmonton AB T6P
1N8, Canada; (780) 440-2131; fax (780) 490-2426; <dmackintosh@canspec.com>.
+
Canspec Group, Inc., 7450 - 18 St. NW, Edmonton AB T6P 1N8, Canada.
Copyright © 2004 by the American Society for Nondestructive Testing, Inc. All
rights reserved.