Introduction
Radiation
detection makes radiography work to give us pictures of the insides
of specimens and allows us to do it safely. Either photographic emulsions
or electronic imaging devices detect penetrating radiation (gammas,
X-rays or neutrons) to form images related to the density thickness
of a specimen. That is, the greater thickness or density portions of
a specimen absorb more radiation than other parts, which then makes
an exposure on the detection system that is a latent image or shadow
picture of the specimen. The latent image on film converts to something
we can see by developing the film.
Radiation safety for radiography
depends greatly upon a combination of common sense and detectors that
provide information about where radiation exists, the quantity of radiation
at each location and the extent of the radiographer's exposure to the
radiation. This short article contains information on the radiation
detectors used in the area of radiation safety.
Knowing
how radiation detectors work can help you use them properly.
Gas Filled Detectors
Gas filled detectors make
up the most often used kinds of radiation detectors. They include Geiger-Muller
(GM) tubes, proportional counters and ion chambers. Generally, they
may all be represented as two electrodes with an applied voltage between
them to collect ions formed by radiation interacting with the gas between
them in the detector. The two electrodes may be flat plates, rings or
a cylinder that forms the outside of the detector and surrounds a center
wire, such as shown in Figure 1.

Figure 1
Diagram of a simple gas filled detector.
If only a small voltage is applied,
then collection of only the ions created by the radiation interacting
with the gas fill occurs (collection of primary ions). Collection of
additional ions begins as the voltage between the electrodes increases.
The additional ions (secondary ions) form when the primary ions accelerate
toward an electrode and collide with the gas molecules that are in their
path. Most ionization occurs when electrons are collected at the center
wire (anode), when the applied voltage is very high and the positive
ions are collected at the outside wall (the cathode) of the detector.
When the volume of gas surrounding the anode is completely ionized,
ion collection for an individual ionizing event or radiation interaction
does not occur until the ions are collected or neutralized. The detector
must collect essentially all of the ions before another can even be
detected. This process limits the rate at which individual events can
be detected.
Figure 2 illustrates the relationship
of ion collection in a gas filled detector versus the applied voltage.
In the ion chamber region, only primary ions are collected. In the proportional
region, primary ions and a number of secondary ions proportional to
the primary ions originally formed are collected. In the GM region,
a maximum number of secondary ions are collected when the gas around
the anode is completely ionized. The detector goes into continuous discharge
if the voltage increases beyond the GM region for a detector and no
individual radiation interaction or events can be detected. Note that
discrimination between kinds of radiation is possible in the ion chamber
and proportional regions but not in the GM region. Each kind of radiation
forms different numbers of primary ions in the detector. However, in
the GM region the number of secondary ions collected per event remains
the same no matter how many primary ions initiate the event. The collection
of the large number of secondary ions in the GM region is known as an
avalanche and produces a large voltage pulse. The large pulse simplifies
the electronics that are necessary for instruments such as survey meters.

Figure 2
Comparison of ion collection versus applied voltage in a gas filled
detector.
Instruments
Collection of only primary ions in the ion chamber region requires expensive,
sensitive electronics but provides information on true radiation exposure.
Such expensive and delicate instruments are used in radiography only
when necessary for accurate radiation exposure values, when multiple
energies of radiation are present such as with X-ray machines and usually
under laboratory conditions but not under field conditions. Proportional
detectors discriminate between types of radiation but require additional
and very stable electronics. Such is the case of detecting neutrons
in radiation fields of both neutron and gamma radiation. Such equipment
is very expensive but worth the cost when required. GM detectors offer
low cost but suffer from a lack of ability to discriminate, slow detection
rates and failure to detect individual events in high radiation fields
(when they become saturated).
Proper design of the GM survey instrument
solves most of the disadvantages and takes advantage of the large electrical
pulses produced per radiation interaction event by keeping the equipment
simple, inexpensive, rugged and dependable. To overcome the lack of
discrimination, the instruments are calibrated for the energy of gamma
radiation similar to that being used by the radiographer. Most often,
gamma radiation from Cs-137 at 0.662 MeV provides the calibration. Only
small errors occur when the radiographer uses Ir-192 (average energy
about 0.34 MeV) or Co-60 (average energy about 1.25 MeV). Use of metal
shields around the GM tubes may help compensate for the differences
in gamma energy of the different radiation sources.
Proper design also helps the GM survey
instrument register radiation fields as low as 2.6 x 10-7
C/kg (1 mR) per hour and as high as 2.6 x 10-4 C/kg (1 R)
per hour as required in the Code of Federal Regulations used
by the Nuclear Regulatory Commission or compliant regulations used in
agreement states. The 2.6 x 10-7 C/kg (1 mR) requirement
allows the instruments to be able to meet regulatory survey requirements
of 5.2 x 10-7 C/kg (2 mR) in any 1 h at barricades. And the
2.6 x 10-4 C/kg (1 R) requirement assures the user that the
detector has not gone into saturation (continuous discharge because
ionizing events are occurring faster than ion collection) in radiation
fields that large. If the detector goes into saturation, the instrument
reads 0, which could be dangerous for a radiographer in a high radiation
field. Again, good design permits the GM detector survey meters to meet
regulatory requirements and still provide ruggedness, simplicity of
use and dependability at a reasonable cost for the radiographer. The
radiographer must, however, turn the survey meter on and use the survey
meter on the job to provide the safety available.
Saturation of GM detectors in high
radiation fields, which prevents ordinary GM tube survey meters from
working, points out another advantage of ion chambers: in high radiation
fields the ion chamber instruments simply read higher or peg at the
highest reading on the meter. Specially designed GM survey meters have
a circuit that pegs the meter when the GM tube goes into saturation
and is unable to produce additional pulses. Most meters simply use a
small tube that does not saturate until the radiation field is greater
than 2.6 x 10-4 C/kg (1 R).
The alarming dosimeters most often
utilize GM detectors for the same reasons that survey meters use them.
However, the pocket dosimeters used by radiographers operate in the
ion chamber region (see Figure 3). Since the pocket dosimeters operate
in the ion chamber region, only primary ions are collected and errors
in correctly reading different radiation energies may be minimized.
The applied voltage makes a small, metalized fiber spring away from
the post where it is attached. The shadow of the fiber is seen on a
screen inside the dosimeter. As the primary ions (formed by the radiation)
neutralize the applied voltage, the fiber moves toward the post and
its shadow moves across a calibrated scale on the screen. A pocket dosimeter
may read a wide range of energies, from X-ray to gamma, correctly while
a GM survey meter cannot because the dosimeter operates in the ion chamber
region and the survey meter operates in the ion avalanche GM region.
Both should be calibrated to be certain they read correctly, and they
should not be used outside of the energy range specified by the manufacturer.

Figure 3
Pocket dosimeter.
Thermoluminescent Dosimeters
Film or thermoluminescent dosimeter badges give radiographers a way
of determining the total radiation exposure for a period of time. Radiation
exposed film turns black when it is developed. Plastic, aluminum, cadmium
and/or lead filters in the holder for the film packet allow determination
of types of radiation and its approximate energy. Lead filters out low
energy radiation so that the film records only high energy radiation
exposure under the lead filters. Plastic filters out most beta particles.
Aluminum and cadmium filter out different energies of X-rays as well
as beta particles. Cadmium converts neutrons into gamma radiation so
that the neutron exposure may be measured. Unfortunately, film also
detects heat and mechanical damage so the film dosimeters must be handled
carefully.
The thermoluminescent dosimeter badges
contain small pieces of material such as lithium fluoride or manganese
sulfate which absorb the radiation in such a fashion that some of the
electrons in the material remain in excited or high energy states for
a long time. When the thermoluminescent dosimeter materials are heated,
they release light in quantities related to their radiation exposure.
The materials can be reused, are relatively rugged and give reliable
information over long periods of time. Also, the badges are generally
smaller than film badges.
Conclusion
Again, none of the radiation safety detectors protect a radiographer
who does not use them or who uses them improperly. Knowing how radiation
detectors work can help you use them properly. While regulations require
their use, only your good sense will make you use them when you are
doing radiography.