Introduction
Magnetism
is everywhere, no matter where you are on this earth and possibly in
the universe. Magnetism originating from within the center of the Earth
permeates your body, your flesh and bones. The first recorded observation
of magnetism dates to AD 1200 when, in Magnesia (a part of Asia Minor),
it was observed that a certain iron ore, called lodestone, was observed
to have peculiar properties, henceforth referred to as magnetism.
Magnetism requires proper instruments
and techniques for its accurate measurement. These are necessary for
both the utilization of magnetism and consideration of the negative
effects that magnetism can produce. Because magnetism is very difficult
to understand and test, we offer some guidelines to aid the technician
in making and recording magnetic measurements.
Source of Magnetism
The very basis of all magnetism is electron spin and rotation, as depicted
in Figure 1. In a free state, atoms have random orientations of electron
spin and rotation. If an object is susceptible to magnetism and is placed
into a magnetic field, its spins and rotations align themselves to be
coherent with the applied field to a limit of intrinsic saturation,
a condition where magnetism increases further only at the permeability
of air.
Magnetic
surveys should be performed at incoming testing and prior to
and during parts assembly.
Direct Component of Magnetism
The direct component of magnetism is either artificially induced
or natural. Intense magnetism in machinery and equipment is produced
by us for our own purposes. Just how much magnetism is required for
a certain purpose and how to measure and control it is of great importance.
When the magnetizing force is removed, the remaining magnetism is referred
to as residual magnetism.

Figure 1 - Schematic
of an atom showing electron spin.

Figure 2 - Characteristic
of a material with high magnetic retentivity. B denotes magnetic
flux density; H denotes magnetizing force.
Magnetic flux density is expressed
in teslas (volt seconds per square meter). The measurement is often
still expressed in gauss (corresponding to the number of lines of magnetic
flux per square centimeter). The magnetizing force is expressed in amperes
(gilberts or ampere turns per centimeter in the centimeter/gram/second
measurement system). Figure 2 shows the character of a material having
high magnetic retentivity (a magnetically hard material). Increasing
magnetizing force H, the magnetic flux density B will
increase along the dashed line of Figure 2, reaching an upper point
a when H reaches value g. Then, as the magnetizing
force is reduced, the flux density B, tending to retain its magnetism,
falls along line d to the residual value Br
as the magnetizing force H reaches zero.
With reverse application of the magnetizing force -H, B
drops to zero when the magnetizing force reaches Hc.
A pattern is followed as it continues, first in the negative direction,
then in the positive direction, to trace out the hysteresis curve of
Figure 2. The remaining residual magnetism could become a nightmare
for the owner or user unless it is removed by downcycling demagnetizing.
An example of this is shown in Figure 3, where progressively smaller
magnetizing forces are applied at alternating polarities, reducing the
residual magnetism to zero.

Figure 3 - Schematic
showing downcycling demagnetization.

Figure 4 - Residual
state br and hc
shown for a material with low magnetic retentivity.

Figure 5 - Schematic
of the earth's magnetic field.
If the part has low magnetic retentivity
(is magnetically soft), its residual state would be br
and hc,
as is depicted in Figure 4, having considerably less magnetism than
the magnetically hard part. Only if the object is very large or long,
such as part of a pipeline, would residual magnetism pose a problem.
Other sources for direct current magnetism are the earth's magnetic
field and objects on the earth, such as lodestone. The earth's magnetic
field performs as if there were a large iron bar extending between the
South and the North Poles, as shown in Figure 5. Measuring low level
magnetic fields in objects can very easily be affected by the earth's
magnetic field, especially near large steel structures.
Measuring the
Direct Component of Magnetism
The generic name for meters used to measure low levels of magnetism
and, notably, the earth's field is magnetometer. The simplest and cheapest
is the dial indicator. Its accuracy is always in question since it can
easily be damaged, either mechanically from being dropped or from exposure
to excessive magnetism. It can measure only fields directed upward into
the case bottom and it cannot be inserted into a gap or crevice where
the most intense fields occur.
An electronic meter (Figure 6) obtains
its direct component of magnetism input signal from a hall sensor. It
is the most universally employed type of sensor for measuring magnetic
fields in the range of 0 to 1 T (0 to 10 000 G). All hall probes should
have at least a 152 mm (6 in.) reach and be durable, with an interconnecting
cable 0.9 to 1.2 m (3 to 4 ft) long to provide access to all areas where
readings are to be made.

Figure 6 - An electronic meter
which obtains its direct component of magnetism through a hall sensor.
Auto ranging of the display range is
necessary for ease of operation and the display should be intuitive
so that field scanning and rough surveying are possible, with no confusion
or digital jump in making readings. The meter should withstand mechanical
shocks and abuse and should perform in the intense fields used in magnetizing
and demagnetizing without damage and with no effect on readings or calibration.
ALTERNATIVE COMPONENT OF MAGNETISM
The alternating component of magnetism exists as a regular and repeatable
pattern of oscillation and is artificially induced. It usually originates
with electric generators and motors and it is strongest in and near
electrical equipment such as power transformers and rotating electrical
machinery. The frequency of alternating magnetism ordinarily bears some
relation to the frequency of electrical power, the speed of rotating
machinery, their harmonics or any combination of the three.
Measuring the Alternating Component
of Magnetism
Tests made on hall sensors prove it to be inaccurate for measuring alternating
component of magnetism fields and combined alternating and direct component
of magnetism fields. Separate alternating and direct component of magnetism
measuring probes, each feeding its normalizing circuitry, constitute
the only accurate way of determining the true magnetic field magnitudes.
MEASURING TECHNIQUES
Accuracy and Precision
There is the mistaken understanding that the more digits in the
display of a digital meter, the greater its accuracy. This is clearly
not the case, as the accuracy of the field measuring probe, its calibration
method and the instrument signal conditioning are what control the meter's
accuracy, regardless of the number of digits in the display.
Polarity Determination
Polarity is the identification of magnetic north and south poles in
the otherwise continuous invisible magnetic field lines, similar to
the earth's magnetic field (Figure
5). Polarity should not be confused with the different magnetizing
field orientations employed during magnetic particle testing - a necessary
technique for locating imperfections in all directions at the surface
and subsurface of an object being tested. Polarity often may not be
important so long as magnetizing is followed by thorough downcycling
demagnetizing using the identical coil setup used to conduct the test.
Polarity is an important factor that must be taken into consideration
when demagnetizing a component of unknown residual magnetism. The technician
must understand this importance and be able to establish reference polarity
if demagnetization is desired. This requires a working knowledge of
the probe and meter to be employed in order to properly note and record
accurate readings, employing a simple notation such as: +x, +yh,
where x is the area reading as positive or a north pole. The
value y is listed when there is a notable maximum reading in
the zone of consideration. The letter h is added next to the
y reading if there is a pronounced peak rather than a general
area maximum value.
Compensating for the Earth's Field
The earth's magnetic field has little direct effect on us or our machinery,
but it does affect the making of low level direct component of magnetism
measurements. The earth's field, normally around 5 x 10-5
T (0.5 G), can reach 1.5 x 10-4
T (1.5 G) or higher, influencing magnetic surveys made in open air,
especially on objects where the residual limit is 2 x 10-4
T (2 G). An object's true residual magnetism can be determined by compensating
for the effect of the earth's field.
Residual magnetism in assembled rotating
machinery should be low enough to prevent the generation of stray voltages
and currents capable of causing damage to bearings or seals with loss
of production and operating efficiency. Magnetism of critical components
when in open air should be reduced to 2 x 10-4
T (2 G) or less.
High Reading Zones
Magnetism is highest near cracks, partings and on outside edges and
corners. Demagnetizing success is achieved when the maximum value meets
the allowable (Figure 7).

Figure 7 - Approximate
reading multipliers.
Low Reading Zones
Low reading zones of magnetism can be deceiving, as they may fail to
reveal high internal magnetism that can be very serious when the part
is assembled or machined (Figure 8). Major causes for this are circular
magnetization and improper welding techniques.

Figure 8 - Low reading zones of
magnetism.
RECOMMENDED FOR MAGNETIC
TESTING
Do not overmagnetize. When magnetizing
for magnetic particle testing, increase the direct current force gradually,
stopping when an optimum indication is obtained, usually around 3 x
10-3
T (30 G). Afterwards, demagnetize using the magnetizing coil. With the
coil still in place, switch to the opposite polarity and set a program
of auto downcycling. Confirm a final magnetism free object using a reliable
meter. Do not employ the circular magnetization head shot technique
for magnetic particle testing, as intense circular magnetism remains,
can generate damaging electrical currents, cannot be measured accurately
and is very difficult to remove. If using alternating current magnetizing
for magnetic particle testing, do not switch on and off but increase
and decrease the magnetizing force gradually.
CONCLUSION
Because magnetism is not visible and can be so damaging if left
behind, an accurate meter should be employed to measure for safe levels
following magnetic particle testing and welding. Magnetic surveys should
be performed at incoming testing and prior to and during parts assembly.
Allowable maximums of residual magnetism for each type of part should
be specified and adhered to. Results of magnetic surveys should be recorded
and maintained for future comparisons, as well as for establishing trends.
* Magnetic Products
and Services, Inc., 2135 Highway 35, Holmdel, NJ 07733; (732) 264-6651;
fax (732) 264-6876; e-mail <support@gaussbusters.com>.
+
Magnetic Products and Services, Inc., 2135 Highway
35, Holmdel, NJ 07733; (732) 264-6651; fax (732) 264-6876.
Copyright ©
2002 by the American Society for Nondestructive Testing, Inc. All rights
reserved.